• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA. BD ANGIOCATH¿ PERIPHERAL VENOUS CATHETER; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA. BD ANGIOCATH¿ PERIPHERAL VENOUS CATHETER; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 381123
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/12/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Event or Problem Description
It was reported while using bd angiocath¿ peripheral venous catheter the needle pierced the catheter.There was no report of patient impact.The following information was provided by the initial reporter: this is a report about needle through catheter.According to the customer's report, when checking the product before use, the hcp found that the needle was piercing through the catheter.
 
Event or Problem Description
It was reported while using bd angiocath¿ peripheral venous catheter the needle pierced the catheter.There was no report of patient impact.The following information was provided by the initial reporter: this is a report about needle through catheter.According to the customer's report, when checking the product before use, the hcp found that the needle was piercing through the catheter.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d9: device available for eval yes, d9: returned to manufacturer on: 01-aug-2022 h6: investigation summary bd received a 22 gauge angiocath device from lot 1239441 for evaluation.A review of the device history record was performed for the reported lot and no quality issues were found during production.Our quality engineer visually inspected the returned unit and observed that the needle had pierced through the catheter tubing.Therefore, based off the visual inspection the engineer was able to verify the reported defect.It was determined that this was likely a manufacturing defect caused by a misalignment between the needle and catheter during the catheter placement process.H3 other text : see h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ANGIOCATH¿ PERIPHERAL VENOUS CATHETER
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA.
ave. pres.juscelino kubitschek
273 francisco bernardino
juiz de fora
Manufacturer (Section G)
BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA.
ave. pres.juscelino kubitschek
273 francisco bernardino
juiz de fora
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15191455
Report Number9610048-2022-00088
Device Sequence Number6014016
Product Code FOZ
UDI-Device Identifier00382903811236
UDI-Public00382903811236
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K151698
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date08/31/2026
Device Catalogue Number381123
Device Lot Number1239441
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/13/2022
Supplement Date Received by Manufacturer08/16/2022
Initial Report FDA Received Date08/09/2022
Supplement Report FDA Received Date09/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-