A review of the device history record has been completed.No deviations or non-conformance's noted.Device evaluation: analysis of the returned device identified: underweight and broken shell.A visual and microscopic analysis was performed which identified: sharp broken on posterior, missing piece of the shell and non-penetrating nicks.A dimension measurement in the shell was performed which identify the thickness within specification.Base on the device analysis the final assessment is: a sharp pattern on posterior of broken device assessed as a surgical impact opening, for the non-penetrating nicks in the shell.Missing shell assessed as inconclusive.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|