• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number TRX-525
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 05/25/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Healthcare professional reported right side exchange from textured to smooth and rupture.The device has been explanted and replaced.
 
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformance's noted.Device evaluation: analysis of the returned device identified: underweight and broken shell.A visual and microscopic analysis was performed which identified: sharp broken on posterior, missing piece of the shell and non-penetrating nicks.A dimension measurement in the shell was performed which identify the thickness within specification.Base on the device analysis the final assessment is: a sharp pattern on posterior of broken device assessed as a surgical impact opening, for the non-penetrating nicks in the shell.Missing shell assessed as inconclusive.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key15192476
Report Number9617229-2022-13524
Device Sequence Number5977857
Product Code FTR
UDI-Device Identifier10888628008557
UDI-Public10888628008557
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/17/2021
Device Model NumberTRX-525
Device Catalogue NumberTRX-525
Device Lot Number2872209
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/29/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/13/2022
Initial Report FDA Received Date08/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/17/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age50 YR
Patient SexFemale
Patient Weight66 KG
-
-