• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 SPIRIT® HYDROCOLLOID ADHESIVE SHEATH; MALE EXTERNAL CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 SPIRIT® HYDROCOLLOID ADHESIVE SHEATH; MALE EXTERNAL CATHETER Back to Search Results
Model Number 35303
Medical Device Problem Code Nonstandard Device (1420)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the if the patient laid there and drink water, it start filling up then the male external catheter would fall off.Patient had been using the products for more than 90 days (batch# jufw0493).Per investigator notification received on 17feb2022.It was stated that the three lots were received for this complaint: (batch# jufu2520), (batch# jufw1132) and (batch# jufv9017) per notification from investigator based on sample evaluation on 26jul2022, it was reported that the male external catheter had strong adhesive (batch# jufw1132).
 
Additional Manufacturer Narrative
The reported event was confirmed as manufacturing related.Although an exact root cause could not be determined a potential root cause could be mechanical failure.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.Labelling review was not required as it would not have prevented the reported event.The actual/suspected device was inspected.
 
Event or Problem Description
It was reported that the if the patient laid there and drink water, it start filling up then the male external catheter would fall off.Patient had been using the products for more than 90 days (batch# jufw0493).Per investigator notification received on 17feb2022.It was stated that the three lots were received for this complaint: (batch# jufu2520), (batch# jufw1132) and (batch# jufv9017) per notification from investigator based on sample evaluation on 26jul2022, it was reported that the male external catheter had strong adhesive (batch# jufw1132).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPIRIT® HYDROCOLLOID ADHESIVE SHEATH
Common Device Name
MALE EXTERNAL CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key15192558
Report Number1018233-2022-06197
Device Sequence Number17543542
Product Code NNX
UDI-Device Identifier00801741070945
UDI-Public(01)00801741070945
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/28/2026
Device Model Number35303
Device Catalogue Number35303
Device Lot NumberJUFW1132
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/26/2022
Supplement Date Received by Manufacturer09/14/2022
Initial Report FDA Received Date08/09/2022
Supplement Report FDA Received Date09/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-