| Model Number |
BEA25-60/I16-40 |
| Medical Device Problem Code |
Separation Problem (4043)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994)
|
| Date of Event |
07/21/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The patient was initially implanted with an afx2 bifurcated stent graft and a vela suprarenal aortic extension to treat an abdominal aortic aneurysm (aaa).Approximately six (6) years post initial procedure, the patient presented with back pain and a type 3a endoleak was identified with no acute rupture.The patient is currently being monitored while an intervention is being discussed.
|
| |
|
Additional Manufacturer Narrative
|
|
The device involved in this event will not be returned for evaluation as it remains implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by the clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx2.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because the device remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the type iiia endoleak with complete component separation is confirmed.The secondary endovascular procedure is confirmed.This is consistent with the reported adverse event/incident.Procedure related harms, device, user, procedure or anatomy relatedness of this complaint could not be determined with the medical records available for review.The final patient status remains unknown.The final patient status was reported pending re-intervention.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx2.H6: health effect - impact code - remove code 4614.
|
| |
|
Event or Problem Description
|
|
The patient was initially implanted with an afx2 bifurcated stent graft and a vela suprarenal aortic extension to treat an abdominal aortic aneurysm (aaa).Approximately six (6) years post initial procedure, the patient presented with back pain and a type 3a endoleak was identified with no acute rupture.The patient is currently being monitored while an intervention is being discussed.After the initial report, additional information was received reporting that reintervention was completed on an unknown date with the implant of a non-endologix long tube stent.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because the device remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the type iiia endoleak with complete component separation is confirmed.This is consistent with the reported adverse event/incident.Procedure related harms, device, user, procedure or anatomy relatedness of this complaint could not be determined with the medical records available for review.The final patient status remains unknown.The final patient status was reported pending re-intervention.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx2.G3: awareness date ¿ updated.H6: investigation finding codes - remove code 3233.H6: investigation conclusion codes - remove code 11.
|
| |
|