| Model Number |
29420 |
| Medical Device Problem Code |
Difficult to Open or Remove Packaging Material (2922)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: the renegade hi flo was returned to boston scientific for analysis.The device was inspected for any damage or irregularities.The renegade was received partially in the shipping hoop.Two fractures were visible before removing from the hoop.They were located 8.7 cm and 44.1 cm from the hub.A kink was noted 46.8 cm from the hub.The shipping hoop was hydrated and the remaining part of the renegade hi flo was removed with no resistance.Because there was no evidence of any product quality deficiencies, it was considered likely that the fractures and kink attributable to handling issues.This concludes the product analysis.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on (b)(6) 2022.It was reported that the microcatheter would not release from the hoop.A 105/10 renegade hi-flo microcatheter was selected for use.During unpacking, it was noted that the microcatheter would not release from the hoop.After flushing the inside and the outside of the microcatheter, it would not release.The procedure was completed with another of the same device.No patient complications were reported.However, device analysis revealed microcatheter fracture located at 44.1cm from the hub.
|
| |
|