Model Number 7170/65
Medical Device Problem Codes Break (1069); Over-Sensing (1438)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/23/2022
Type of Reportable Event Malfunction
Event or Problem Description
Related manufacturer report number: 2017865-2022-17156.It was reported that noise, oversensing and suspected lead damage was observed on the right ventricular (rv) lead and myopotential oversensing was observed on the device.No intervention has been performed to resolve the event at this time.The patient was in stable condition and will continue to be monitored.
 
Event or Problem Description
Additional information received indicated the lead was capped and replaced to resolve the event.The patient was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURATA STS OPTIM PASSIVE FIXATION
Common Device Name
DEFIBRILLATION LEAD
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key15192827
Report Number2017865-2022-17157
Device Sequence Number14844503
Product Code LWS
UDI-Device Identifier05414734502436
UDI-Public05414734502436
Combination Product (Y/N)Y
Initial Reporter CountryIT
PMA/510(K) Number
P950022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/15/2011
Device Model Number7170/65
Device Catalogue Number7170-65
Device Lot Number2997476
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/25/2022
Supplement Date Received by Manufacturer01/13/2023
Initial Report FDA Received Date08/09/2022
Supplement Report FDA Received Date02/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/15/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-