| Model Number |
84591 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/15/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a shaft break occurred.A 130cm direxion hi-flo fathom-16 system was selected for use.During the procedure, it was noted that the catheter had a break along the shaft which caused it to be unusable.There were no complications reported and the patient is doing fine after the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by manufacturer: the device was returned for analysis.The device shaft was microscopically analyzed for any damage.The devices shaft showed 1 fracture located 52.5cm from the hub.The inner liner showed a kink.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.The fractured shaft was confirmed.
|
| |
|
Event or Problem Description
|
|
It was reported that a shaft break occurred.A 130cm direxion hi-flo fathom-16 system was selected for use.During the procedure, it was noted that the catheter had a break along the shaft which caused it to be unusable.There were no complications reported and the patient is doing fine after the procedure.
|
| |
|