|
Product complaint # (b)(4).Investigation findings: (b)(4) ¿ photo evaluation.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Photo investigation summary: investigation summary: this is an analysis of a photo submitted for evaluation.During the visual analysis the following was observed: the photo shows one aluminum package, product code vcp426.The image is not clear enough to determine failure mode or reported condition.Based on the set of photos review, the event describe is not confirmed, however no conclusion or root cause could be determined.This is a combination product, and the event has been reviewed for both the suture and the triclosan.Trade name - irgacare®.Active ingredient(s) ¿ triclosan.Dosage form ¿ suture/solid/parenteral.Strength ¿ = 275 gm.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Event related to mw # 2210968-2022-06403 and mw # 2210968-2022-06404.This report is being submitted pursuant to the provisions of 21 cfr part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|