BOSTON SCIENTIFIC CORPORATION LEVEEN SUPERSLIM; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 36705
Medical Device Problem Code Failure to Unfold or Unwrap (1669)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the needle was difficult to extend.A leveen needle electrode 2.0/17/15 was selected for use in a radiofrequency ablation procedure, to treat the liver.During the procedure, a needle was inserted and attempted to deploy close to the lesion.However, the handle was too tight to be used, even though it was tried again.The procedure was completed with another of the same device.The patient was in stable condition and no complications occurred.
 
Additional Manufacturer Narrative
Device eval by manufacturer: the device was returned with its respective radio frequency ablation electrode cable, and no damages were observed.Functional inspection was performed, and the handle was extended and retracted, and the needle/tines could be activated without any problems.
 
Event or Problem Description
It was reported that the needle was difficult to extend.A leveen needle electrode 2.0/17/15 was selected for use in a radiofrequency ablation procedure, to treat the liver.During the procedure, a needle was inserted and attempted to deploy close to the lesion.However, the handle was too tight to be used, even though it was tried again.The procedure was completed with another of the same device.The patient was in stable condition and no complications occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEEN SUPERSLIM
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15195391
Report Number2124215-2022-29695
Device Sequence Number17636662
Product Code GEI
UDI-Device Identifier08714729441694
UDI-Public08714729441694
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K113090
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/17/2024
Device Model Number36705
Device Catalogue Number36705
Device Lot Number0028409814
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/18/2022
Supplement Date Received by Manufacturer01/09/2023
Initial Report FDA Received Date08/09/2022
Supplement Report FDA Received Date01/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient Age35 YR
Patient SexFemale
-
-