| Model Number |
TRM-405 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
05/02/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).A review of the device history record has been completed.No deviations or non-conformance's noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.The reason for reoperation is: "left implant rupture found on ct scan".
|
| |
|
Event or Problem Description
|
|
Healthcare provider reported a left side rupture found on ct scan and textured implant recall.The device has been explanted and replaced.
|
| |
|
Event or Problem Description
|
|
Healthcare provider reported a left side rupture found on ct scan and textured implant recall.The device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: rupture: was observed, broken device on posterior side assessed, as surgical impact opening (shell thickness within specification).Anxiety-product/procedure: unable to observe, as it is a medical event.That is not related to the device.Additional observations: black particles on the shell are observed.Non penetrating nick is observed, assessed as non-penetrating nick.No further actions are required, as the device was implanted.
|
| |
|
Event or Problem Description
|
|
Healthcare provider reported, a left side rupture, found on ct scan.And textured implant recall.The device has been explanted and replaced.
|
| |
|
Search Alerts/Recalls
|