• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHT MEDICAL TECHNOLOGY INC RESECTION ADJUSTMENT BLOCK INFINITY TOTAL ANKLE SYSTEM; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHT MEDICAL TECHNOLOGY INC RESECTION ADJUSTMENT BLOCK INFINITY TOTAL ANKLE SYSTEM; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 33600030
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/14/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Analysis results are not available at the time of this report.A follow-up report will be sent when the analysis is complete.
 
Event or Problem Description
It was noticed the resection adjustment block wouldn't translate proximal distal at the end of the procedure.The purple knob is locked and won¿t really turn with significant force applied.
 
Additional Manufacturer Narrative
The reported event could not be confirmed, since the returned device is conforming to specifications and fully functional.The device inspection revealed the following: examination of the returned device finds normal wear and tear typical for a reusable device that has been in service for years.There is damage/deformation to the lockout screws and the recesses of the adjustment knobs.Examination of the proximal/distal knob, shaft thread and spring did not find any notable damage or obstructions.Attempts to rotate the proximal/distal knob found the knob jammed in the returned state.Lubricant was applied to the threading and recesses of the knob and threaded shaft.Force was applied to the knob and the jam released.The knob was able to move from the most proximal and distal positions without obstruction.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.The instructions for use instructs user that: ¿use medical devices in accordance with their labeled indications and wright¿s instructions for use, especially during insertion and removal.Inspect devices prior to use for damage during shipment or storage or any out-of-box defects that might increase the likelihood of fragmentation during a procedure.Inspect devices immediately upon removal from the patient for any signs of breakage or fragmentation.If the device is damaged, retain it to assist with wright¿s analysis of the event.For devices with hinged/mated surfaces or moving components, a biocompatible, surgical grade lubricant intended for heat sterilized medical instruments (miltex® spray lubrication or equivalent) should be used per the manufacturer¿s guidelines.Apply spray lubricant to the moving components and/or hinged/mated components where appropriate prior to the sterilization process.¿.Based on investigation, the root cause was attributed to a user related issue i.E.Customer-improper preparation of (re)use.The failure was caused due to not applying enough lubricant to the device resulting in jamming.If any further information is provided, the complaint report will be updated.
 
Event or Problem Description
It was noticed the resection adjustment block wouldn't translate proximal - distal at the end of the procedure.The purple knob is locked and won¿t really turn with significant force applied.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESECTION ADJUSTMENT BLOCK INFINITY TOTAL ANKLE SYSTEM
Common Device Name
PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY INC
1023 cherry rd
memphis TN 38117
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15195593
Report Number3010667733-2022-00268
Device Sequence Number17534373
Product Code HSN
UDI-Device Identifier00889797004008
UDI-Public00889797004008
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K123954
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number33600030
Device Catalogue Number33600030
Device Lot Number1333532
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/14/2022
Supplement Date Received by Manufacturer11/11/2022
Initial Report FDA Received Date08/09/2022
Supplement Report FDA Received Date12/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/21/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-