| Model Number |
00884450009390 |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/29/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The suspect device is not expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The account alleges that the catheter broke into two parts inside the patient during endovascular coil aneurysm repair.Patients had tortuous right external iliac arteries.The foreign body was removed from this patient with a endovascular snare.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device was not returned for evaluation.A photograph was provided by the account.The complaint is confirmed but the root cause could not be determined.The device history record was reviewed, and no exception documents were found.A search of the complaint database was performed and no similar complaints for this lot number were identified.Should the device be returned later, the investigation will be reopened.
|
| |
|