| Model Number |
VICM5_13.7 |
| Medical Device Problem Codes |
Misfocusing (1401); Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Codes |
Intraocular Pressure Increased (1937); Blurred Vision (2137); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
07/08/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Claim # (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated the surgeon implanted a 13.7mm vicm5_13.7 implantable collamer lens, -10.50 diopter, into the patients right eye (od) on (b)(6) 2022.The lens was explanted on (b)(6) 2022 due to excessive vaulting, significant reduction of irido-corneal angles and blurred vision.The lens was replaced with a shorter lens and the problem was resolved.Medication was administered.The lens was discarded.Additional information has been requested but none has been forthcoming.
|
| |
|
Additional Manufacturer Narrative
|
|
B5: there was high pressure, zonular instability, iris prolapse, iris atrophy and shallowing of ac.H6: health impact- clinical code: 4581 - zonular instability, iris prolapse, iris atophy, shallowing of a/c.H6: health effect impact code: 4644 - topical and oral anti-ocular hypertensive medication.Claim # (b)(4).
|
| |
|
Search Alerts/Recalls
|