| Model Number |
M651G |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/26/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc., or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts are being made to obtain the following information. to date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Could you please clarify if the suture & the needle got broken? could you please clarify if the suture was separated from the needle in the swage part? device return follow up.Associated events submitted via mw# 2210968-2022-06427.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a cabg procedure on (b)(6) 2022 and the suture was used.During surgery, the needle and wire are continuously broken when suturing in sternum closure.The operator tried to maintain proper tension for needle penetration.There were no adverse patient consequences.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 9/15/2022.H6 component code: g07002 no device problem found.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that it was received six needle-suture that pertain to the product code m651g.In order to evaluate the condition of the returned samples the swage and attachment area was noted to be as expected.The sutures were examined along of the strand and no issue related to breakage suture were observed during evaluation.In additionally, the samples were tested by resistance test to needle and met the requirements.The functional test was performed using an instron equipment and the tensile force was above the minimum requirements.As part of ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|