Model Number BN7TCFJ4L
Medical Device Problem Codes Material Protrusion/Extrusion (2979); Patient Device Interaction Problem (4001)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/14/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Since no device has been received for analysis, no product investigation can be performed, and the customer complaint cannot be confirmed.A device history record evaluation was performed for the finished device (b)(4) number, and no interal action related to the complaint was found during the review.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.This report is being submitted pursuant to the provisions of 21 cfr, part 4.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number (b)(4).
 
Event or Problem Description
It was reported that a patient underwent an atrioventricular nodal reentry tachycardia (avnrt) ablation procedure with an ez steer¿ nav bi-directional electrophysiology catheter and an electrode damaged without foreign material ¿ sharp rough edges issue occurred.A steam pop was heard in the middle of the procedure.The catheter was removed from the body and they noticed the inner electrode had popped out of the blue lumen.Nothing else had fallen out of the catheter and the electrode was still attached.The catheter was replaced with a smarttouch sf catheter and the issue resolved.Additional information was received on the event.The damage was to the electrode, there were no wires that the biosense webster, inc.Representative could see.The electrode seemed like smooth edges.There was no resistance or difficulty during insertion or removal of the device.The electrode stayed attached within the lumen just a hole was created with part of the electrode exposed.The electrode was visible.No sheath was used.No medical intervention was required.The patient has not exhibited any neurological symptoms since the procedure was completed.Based on the available information, this event was assessed as mdr reportable for an ¿electrode damaged without foreign material ¿ sharp rough edges¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EZ STEER¿ NAV BI-DIRECTIONAL ELECTROPHYSIOLOGY CATHETER
Common Device Name
ELECTRODE, PERCUTANEOUS, CONDUCTION TISSUE ABLATION
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez
MX  
Manufacturer Contact
michelleann garcia
31 technology dr
irvine, CA 92618
646591-7981
MDR Report Key15196559
Report Number2029046-2022-01834
Device Sequence Number20393711
Product Code LPB
UDI-Device Identifier10846835003031
UDI-Public10846835003031
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P990025/S12
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/23/2022
Device Model NumberBN7TCFJ4L
Device Catalogue NumberBN7TCFJ4L
Device Lot Number30311247M
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/14/2022
Initial Report FDA Received Date08/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/24/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SMARTTOUCH SF CATHETER.
Patient SexUnknown
-
-