| Model Number |
MD-151 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/07/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported, the halogen lamp ( md-151 ) did not light when the facility staff replaced it as it was out on the clh-sc (halogen light source).Another new md-151 was used and was noted that it worked without any issues and had no effect on the case.The issue occurred at preparation for use.There was no patient harm, no user injury reported due to the event.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(6).The subject device was evaluated.Device evaluation and inspection, service repair noted the following findings: return device: serial is unknown.Visual inspection: there was something wrong with the appearance.Actual machine confirmation (functional testing ): the defect has been reproduced.(filament is broken) operation check: when connected and operated according to the instruction manual, there are no abnormalities other than the defect found (broken filament).Investigation is ongoing.This report will be supplemented accordingly following investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device history record was unable to be reviewed for this device since the serial number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, the event was caused by a broken filament, but the root cause of the failure cannot be identified.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|