• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT CONFIRM; IMPLANTABLE CARDIAC MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT CONFIRM; IMPLANTABLE CARDIAC MONITOR Back to Search Results
Model Number DM2102
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/20/2022
Event Type  malfunction  
Event Description
During device explant due to normal battery depletion, detachment of the device header was observed.The event was resolved by explanting the device.The patient was in stable condition.
 
Manufacturer Narrative
Customer complaint of detachment of device header was confirmed.As received, device was visually inspected, and header of the device was found to be cracked and separated from the can due to excessive force by the user.This is consistent with improper handling of the device header by user.Device was received in backup mode with battery at end of service (eos) level due to normal usage.The implant duration exceeds the projected longevity indicating normal battery depletion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONFIRM
Type of Device
IMPLANTABLE CARDIAC MONITOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key15198443
MDR Text Key304968800
Report Number2017865-2022-17251
Device Sequence Number1
Product Code MXC
UDI-Device Identifier05414734502573
UDI-Public05414734502573
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K133481
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2019
Device Model NumberDM2102
Device Catalogue NumberDM2102
Device Lot NumberS000064645
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/27/2022
Initial Date FDA Received08/10/2022
Supplement Dates Manufacturer Received10/28/2022
Supplement Dates FDA Received11/02/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/06/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-