OSCOR INC. SAFESHEATH II 9F; INTRODUCER, CATHETER Back to Search Results
Model Number SS9
Medical Device Problem Code Material Too Rigid or Stiff (1544)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/11/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the introducer did not peel completely.Physician had to use a scalpel to remove part that wasn't peeled.The patient a 75 year old female was not harmed.
 
Additional Manufacturer Narrative
Conclusion not yet available, evaluation in process.A follow-up report will be submitted as soon as the investigation is complete.Oscor inc.Is submitting the above report to comply with 21 cfr part 803, the medical device reporting regulation.The report is based upon information obtained by oscor inc.Which the company may not have been able to investigate or verify prior to the date the report was required by fda.This report does not constitute an admission that the device, oscor inc., or its employees, caused or contributed to the event described in this report.Nor does this report reflect a conclusion by fda, oscor inc., or its employees, that the report constitutes an admission that the device, oscor inc., or its employees caused or contributed to the event described in this report.
 
Additional Manufacturer Narrative
The following section was corrected: g4: k122084.The device was used in treatment.The device was discarded and not returned for analysis.Therefore, the clinical observation could not be confirmed.However, the following controls are in place to mitigate the reported product issue.The device history records were reviewed to confirm that the device passed all applicable in-process and final inspections.Per procedure, adelante-s introducer sheath in process and final inspections destructive testing sampling plan ansi z 1.4, special level 4, aql 0.40 reduced break and peel test using samples from fit check as described in procedure.Manually break the sheath and hub and verify that the seal splits easily without extreme elongation.Also verify that the split cap and seal remain secure and do not break free or loosen from the sheath hub.Manually peel the sheath and verify the sheath peels easily along the sheath body and tip.The instructions for use (ifu) adelante safesheath ii, informs the user: flush sheath with 5cc of saline immediately before peeling sheath away in order to minimize back bleeding.Withdraw sheath and valve over the lead or catheter and from the vessel, while keeping the lead in place.Sharply snap the tabs of valve housing in a plane perpendicular to the long axis of the sheath to split the valve and peel sheath apart while withdrawing from the vessel.Based on the investigation, a capa is not required as findings did not identify a design, labeling or manufacturing non-conformity.In addition, there was no new failure mode identified and the risk remains acceptable.The event will be re-evaluated if additional information becomes available.Oscor will continue to monitor this event type and risk.Oscor inc.Is submitting the above report to comply with 21 cfr part 803, the medical device reporting regulation.The report is based upon information obtained by oscor inc.Which the company may not have been able to investigate or verify prior to the date the report was required by fda.This report does not constitute an admission that the device, oscor inc., or its employees, caused or contributed to the event described in this report.Nor does this report reflect a conclusion by fda, oscor inc., or its employees, that the report constitutes an admission that the device, oscor inc., or its employees caused or contributed to the event described in this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFESHEATH II 9F
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
OSCOR INC.
3816 desoto blvd.
palm harbor FL 34683 1816
Manufacturer (Section G)
OSCOR INC.
3816 desoto blvd.
palm harbor FL 34683 1816
Manufacturer Contact
daniel naut
3816 desoto blvd.
palm harbor, FL 34683-1816
7279352511
MDR Report Key15198510
Report Number1035166-2022-00095
Device Sequence Number19360813
Product Code DYB
UDI-Device Identifier30891492002259
UDI-Public30891492002259
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K122084
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSS9
Device Catalogue NumberSS9
Device Lot NumberDP-14348
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/18/2022
Supplement Date Received by Manufacturer12/30/2022
Initial Report FDA Received Date08/10/2022
Supplement Report FDA Received Date01/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age75 YR
Patient SexFemale
-
-