Model Number M0066703011
Medical Device Problem Codes Break (1069); Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/08/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a device analysis could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a sensor wire was used during a mastectomy with portacath placement procedure performed on (b)(6) 2022.During the procedure, upon removal of the sensor wire, the corewire broke.It remained in the internal jugular vein and was picked up on the radiology exams.This required a return to surgery for removal.The fragment was removed intact and was approximately two inches in length.The patient was undergoing another type of surgery in addition to this is portacath placement.Per the ifu the sensor urological guidewire is intended to facilitate the placement of endourological instruments during diagnostic or interventional procedures.These guidewires are not intended for coronary artery, vascular or neurological use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSOR
Common Device Name
STYLET FOR CATHETER, GASTRO-UROLOGY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15198590
Report Number3005099803-2022-04367
Device Sequence Number2162926
Product Code EZB
UDI-Device Identifier08714729302612
UDI-Public08714729302612
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Risk Manager
Type of Report Initial
Report Date (Section B) 08/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/10/2023
Device Model NumberM0066703011
Device Catalogue Number670-301
Device Lot Number0025839986
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/13/2022
Initial Report FDA Received Date08/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/10/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age46 YR
Patient SexFemale
Patient Weight56 KG
Patient RaceWhite
-
-