Model Number NV GXL LNR, LIPPED, 32MM ID GROUP 4 CUPS
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Naturally Worn (2988)
Health Effect - Clinical Codes Failure of Implant (1924); Joint Dislocation (2374)
Date of Event 07/18/2022
Type of Reportable Event Serious Injury
Event or Problem Description
As reported, approximately 6.5 years post initial left tha, surgeon made decision to revise hip due to patient dislocating posteriorly multiple times.Patient had a 60mm novation crown cup cluster-hole shell group, a novation lipped liner group 4 32mm, and a 32mm +0 cocr head on a size 10 novation element ha coated press fit stem.Surgeon revised patient via poly swap by going posterior approach and found the lipped liner faced anterior.Surgeon observed that there was some advanced wear within the poly and the head appeared to be moving around within the poly abnormally as it wasn't held within the liner as normal.Surgeon removed liner and removed head.Surgeon switched to a ceramic 36 option head with a +3.5 sleeve and went with a novation xle 10 degree face changing liner group 4 36mm.
 
Additional Manufacturer Narrative
Concomitant medical products: cocr fem head 32mm +0 offset 12/14 (cat# 142-32-00 / serial# (b)(4)).Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Additional Manufacturer Narrative
After further review of additional information received the following sections g3, g6, h2, h3, h6, h7 and h9 have been updated accordingly.(h3 the revision reported was likely the result of the orientation of the acetabular cup/liner, edge loading/impingement, patient conditions, or a combination of the three, which led to instability, recurrent dislocation and increased prosthesis wear.However, this cannot be confirmed as the explants were not available for evaluation.The most probable root cause associated with the reported event of "prosthesis wear¿ is associated with material damage to a surface, usually involving progressive loss or displacement of material, due to relative motion between that surface and a contacting substance or substances.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay corrected information.The following sections were corrected: d8, h6: clinical codes, medical device problem codes, component code, investigation conclusion.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NV GXL LNR, LIPPED, 32MM ID GROUP 4 CUPS
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
kate jacobson
2320 nw 66 court
gainesville, FL 32653
3523771140
MDR Report Key15198652
Report Number1038671-2022-00906
Device Sequence Number6451693
Product Code JDI
UDI-Device Identifier10885862023339
UDI-Public10885862023339
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K070479
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/11/2019
Device Model NumberNV GXL LNR, LIPPED, 32MM ID GROUP 4 CUPS
Device Catalogue Number132-32-54
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/21/2022
Supplement Date Received by Manufacturer12/15/2022
12/05/2025
Initial Report FDA Received Date08/10/2022
Supplement Report FDA Received Date01/10/2023
12/11/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/13/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1729-2022
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
Patient SexFemale
-
-