Model Number 16-02-85 |
Device Problem
Temperature Problem (3022)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/19/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in san diego, california.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova deutschland received a report that a heater-cooler system 3t displayed an error message during procedure.The affected circuit was not specified by the customer and a malfunction of the stirrer motor on the patient side cannot be excluded.Thus, the event is conservatively considered as reportable.There was no patient injury.
|
|
Manufacturer Narrative
|
H.10: a livanova field service representative was dispatched to the facility to troubleshoot the reported malfunction.The service representative was able to confirm the reported issue since the stirrer motor was found blocked.The stirrer motor was replaced in order to fix the issue.A functional control and a test run were performed and passed with no further issue shown.The device was returned to service.The most likely root cause of the reported event is related to mechanical/electronic failure of the stirrer motor.
|
|
Event Description
|
See initial report.
|
|