LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-85
Device Problem Temperature Problem (3022)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/19/2022
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in san diego, california.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova deutschland received a report that a heater-cooler system 3t displayed an error message during procedure.The affected circuit was not specified by the customer and a malfunction of the stirrer motor on the patient side cannot be excluded.Thus, the event is conservatively considered as reportable.There was no patient injury.
 
Manufacturer Narrative
H.10: a livanova field service representative was dispatched to the facility to troubleshoot the reported malfunction.The service representative was able to confirm the reported issue since the stirrer motor was found blocked.The stirrer motor was replaced in order to fix the issue.A functional control and a test run were performed and passed with no further issue shown.The device was returned to service.The most likely root cause of the reported event is related to mechanical/electronic failure of the stirrer motor.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key15198672
MDR Text Key298045896
Report Number9611109-2022-00408
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 08/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-85
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/19/2022
Initial Date FDA Received08/10/2022
Supplement Dates Manufacturer Received08/22/2022
Supplement Dates FDA Received09/20/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/28/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-