MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION GLADIATOR ELITE; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION GLADIATOR ELITE; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 24686
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/04/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The stenosed target lesion was located in the left upper extremity.A 7.0 x40, 40cm gladiator elite balloon catheter was advanced for dilation.The balloon was inflated first at 20 atmospheres (atm) but the stenosis was not resolved.When it was inflated for the second time at 18 atm, the balloon burst.The device was removed and the procedure was completed with another of the same device.No patient complications or injuries reported and the patient was stable post-procedure.
 
Event or Problem Description
It was reported that balloon rupture occurred.The stenosed target lesion was located in the left upper extremity.A 7.0 x40, 40cm gladiator elite balloon catheter was advanced for dilation.The balloon was inflated first at 20 atmospheres (atm) but the stenosis was not resolved.When it was inflated for the second time at 18 atm, the balloon burst.The device was removed and the procedure was completed with another of the same device.No patient complications or injuries reported and the patient was stable post-procedure.It was further reported that the lesion was 80% stenosed.
 
Additional Manufacturer Narrative
Device evaluated by mfr: the device was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.A longitudinal tear was identified in the balloon material.The tear measured 53 mm in length and extended from a position 6mm proximal of the proximal markerband to 5mm distal of the distal markerband.An examination of the balloon material and markerbands identified no issues which could potentially have contributed to this complaint.A visual examination observed no damage to the tip of the device.A visual and tactile examination found no kinks or damage to the shaft of the device.Both markerbands were undamaged and present on the shaft of the device.This concludes the product analysis.
 
Event or Problem Description
It was reported that balloon rupture occurred.The stenosed target lesion was located in the left upper extremity.A 7.0 x40, 40cm gladiator elite balloon catheter was advanced for dilation.The balloon was inflated first at 20 atmospheres (atm) but the stenosis was not resolved.When it was inflated for the second time at 18 atm, the balloon burst.The device was removed and the procedure was completed with another of the same device.No patient complications or injuries reported and the patient was stable post-procedure.It was further reported that the lesion was 80% stenosed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLADIATOR ELITE
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
pmt 741 persiaran cassia selat
bandarcassia, pulau pinan 14110
MY   14110
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15198800
Report Number2124215-2022-29804
Device Sequence Number2252749
Product Code LIT
UDI-Device Identifier08714729809784
UDI-Public08714729809784
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K132810
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/08/2024
Device Model Number24686
Device Catalogue Number24686
Device Lot Number0027800529
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/05/2022
Supplement Date Received by Manufacturer08/05/2022
09/01/2022
Initial Report FDA Received Date08/10/2022
Supplement Report FDA Received Date08/16/2022
09/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age66 YR
Patient SexFemale
Patient Weight58 KG
-
-