• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. ELECTROSURGICAL UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. ELECTROSURGICAL UNIT Back to Search Results
Model Number UES-40
Medical Device Problem Code Output above Specifications (1432)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device referenced in this report has not been returned to olympus for evaluation.The investigation is ongoing.If additional information becomes available, this report will be supplemented accordingly.
 
Event or Problem Description
The customer reported to olympus, the power values were above those recommended on the electrosurgical unit.There were no reports of patient harm associated with this event.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Approximately over 11 years have passed since the device was manufactured.Based on the results of the legal manufacturer's investigation, a probable cause could not be identified due to the product not being returned and no additional information was obtained.A definitive root cause cannot be identified.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECTROSURGICAL UNIT
Common Device Name
ELECTROSURGICAL UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer Contact
masaharu hirose
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8-061
JA   961-8061
426422891
MDR Report Key15198844
Report Number3002808148-2022-00824
Device Sequence Number2162435
Product Code GEI
UDI-Device Identifier04953170287275
UDI-Public04953170287275
Combination Product (Y/N)N
Initial Reporter CountryPO
PMA/510(K) Number
K030194
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberUES-40
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/14/2022
Supplement Date Received by Manufacturer10/20/2022
Initial Report FDA Received Date08/10/2022
Supplement Report FDA Received Date11/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/22/2011
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1