STERIS MEXICO, S. DE R.L. DE C.V. EVOLUTION TRANSFER CARRIAGE Back to Search Results
Model Number AX083020003
Medical Device Problem Code Device Tipped Over (2589)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
The user facility reported that their loading cart fell to the floor while an employee was gliding it onto their 66" transfer carriage.No report of injury.
 
Additional Manufacturer Narrative
A steris service technician arrived onsite and observed that one of the racks from the loading cart was damaged.When the loading cart fell to the floor, it made contact with the sterilizer's door trim subsequently causing the damage to occur.The technician inspected the transfer carriage and found the height on the engagement rollers required adjustment.The loading cart was not properly locking to the transfer carriage subsequently causing the loading cart to fall.The technician made adjustments to the engagement rollers, tested the transfer carriage and confirmed the equipment to be operating properly.The transfer carriage and sterilizer were returned to service.No additional issues have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUTION TRANSFER CARRIAGE
Common Device Name
TRANSFER CARRIAGE
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX  67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX   67190
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key15199123
Report Number3005899764-2022-00036
Device Sequence Number2162445
Product Code FLE
UDI-Device Identifier00724995150389
UDI-Public00724995150389
Combination Product (Y/N)N
Initial Reporter CountryCA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberAX083020003
Device Catalogue NumberAX083020003
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/14/2022
Initial Report FDA Received Date08/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/27/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-