| Model Number |
212723 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/05/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: udi: (b)(4).Incomplete.Initial reporter occupation: reporter is a j&j sales representative.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported by the sales rep in united kingdom that when a patient underwent an mri scan on an unknown date, something ping in the shoulder.According to the report, ct/x-rays revealed that the bioknotless br anchor w/ panacryl device appeared to have become detached.It was further reported that it revealed what looks like to be a metal component (or very dense area) associated with the device.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the complaint device is not being returned, it was implanted in the patient, therefore unavailable for a physical evaluation.No lot numbers were supplied which precludes conducting a dhr review or a lot of specific search in the complaints handling system.As part of depuy mitek¿s quality process all devices are manufactured, inspected, and released to approved specifications.With the information provided, and without the complaint device to evaluate, we cannot determine a root cause for the reported failure.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary = > the complaint device is not being returned, it was implanted in the patient, therefore unavailable for a physical evaluation.A manufacturing record evaluation was performed for the finished device lot number and no non-conformance was identified.As part of depuy mitek¿s quality process all devices are manufactured, inspected, and released to approved specifications.With the information provided, and without the complaint device to evaluate, we cannot determine a root cause for the reported failure.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: b3, b5: subsequent follow-up with the customer, additional information was received regarding the event.Therefore, the event date and event description have been updated accordingly.D4, h4: the lot number, expiration date and device manufacture date were reported as unknown on the initial report; and have been updated accordingly.Therefore, udi: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported by the sales rep in united kingdom that when a patient underwent an mri scan on (b)(6) 2021, something ping in the shoulder.According to the report, ct/x-rays revealed that the bioknotless br anchor w/ panacryl device appeared to have become detached.It was further reported that it revealed what looks like to be a metal component (or very dense area) associated with the device.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
|
| |
|