• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GYRUS ACMI L.P. SHOCKPULSE LITHOTRIPSY TRANSDUCER.; LITHOTRIPTOR, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GYRUS ACMI L.P. SHOCKPULSE LITHOTRIPSY TRANSDUCER.; LITHOTRIPTOR, ULTRASONIC Back to Search Results
Model Number SPL-T
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported that, during preparation for use, the suction connector on the subject device was found to be damaged.The device was sent to an olympus service center for evaluation.During inspection and testing, service found the tip of the suction port was broken and damaged.This report is being submitted for the malfunction found during evaluation (broken suction port).There was no effect on the patient due to the event.
 
Additional Manufacturer Narrative
During inspection and testing, service found the tip of the suction port was broken and damaged.A review of the device history record found no deviations that could have caused or contributed to the broken port.The device shipped according to specifications.Based on the results of the legal manufacturer's investigation, it is likely the broken port was due to improper handling of the device.However, a definitive root cause of the reported issue could not be identified.The device's instruction manual provides the following, which may help to prevent the issue: "this product is a precision device; handle it with care.Avoid rough or violent handling, which may cause equipment damage." olympus will continue to monitor field performance for this device.Investigation activities have been opened to manage the actions related to this issue and any required mdr reporting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHOCKPULSE LITHOTRIPSY TRANSDUCER.
Common Device Name
LITHOTRIPTOR, ULTRASONIC
Manufacturer (Section D)
GYRUS ACMI L.P.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer Contact
todd brill
9600 louisiana avenue north
brooklyn park, MN 55445
5082077661
MDR Report Key15200081
Report Number3005975494-2022-00026
Device Sequence Number2111638
Product Code FEO
UDI-Device Identifier00855279005023
UDI-Public00855279005023
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K142428
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSPL-T
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/25/2020
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/20/2020
Initial Report FDA Received Date08/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/21/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-