ARTHREX, INC. LEFTY TIGHT LAG SCREW, Ø10.5MM X 90MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
|
Back to Search Results |
|
| Model Number |
LEFTY TIGHT LAG SCREW, Ø10.5MM X 90MM |
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
11/22/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2021, it was reported by a sales representative via email that an 1192-090 lefty tight lag screw broke.This was discovered during a procedure on (b)(6) 2021.When the surgeon attempted to advance the screw, it broke inside the patient.The screw was retrieved and was replaced with an 1099-090 galileo lag screw.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint confirmed.Upon visual inspection, it was noted that the second thread was broken off.The broken piece was not returned for evaluation.The cause of this remains undetermined.
|
| |
|
Search Alerts/Recalls
|
|
|