• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONCORD MANUFACTURING DRY ACID 99 GALLON UNIT, ASSY TO SHIP; DIALYSATE CONCENTRATE FOR HEMODIALYSIS (LIQUID OR POWDER)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONCORD MANUFACTURING DRY ACID 99 GALLON UNIT, ASSY TO SHIP; DIALYSATE CONCENTRATE FOR HEMODIALYSIS (LIQUID OR POWDER) Back to Search Results
Model Number 160148
Medical Device Problem Code Smoking (1585)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
A biomedical technician (biomed) at a user facility reported that smoke was coming from a fresenius dry acid mixer.The machine was removed from service.The customer requested an onsite evaluation by a fresenius field service technician (fst) however this request was cancelled the following day.There was no reported harm to any patients or individuals because of this malfunction.Additional information was requested, however a response has not been received.No parts were returned to the manufacturer for physical evaluation.
 
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Additional Manufacturer Narrative
Plant investigation: no parts were returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation was performed by a fresenius field service technician (fst).A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The investigation y the manufacturer did not determines that there was a causal relationship between the objective evidence and the alleged event; the alleged event is unconfirmed.
 
Event or Problem Description
A biomedical technician (biomed) at a user facility reported that smoke was coming from a fresenius dry acid mixer.The machine was removed from service.The customer requested an onsite evaluation by a fresenius field service technician (fst) however this request was cancelled the following day.There was no reported harm to any patients or individuals because of this malfunction.Additional information was requested, however a response has not been received.No parts were returned to the manufacturer for physical evaluation.
 
Event or Problem Description
A biomedical technician (biomed) at a user facility reported that smoke was coming from a fresenius dry acid mixer.The machine was removed from service.The customer requested an onsite evaluation by a fresenius field service technician (fst) however this request was cancelled the following day.There was no reported harm to any patients or individuals because of this malfunction.Additional information was requested, however a response has not been received.No parts were returned to the manufacturer for physical evaluation.
 
Event or Problem Description
A biomedical technician (biomed) at a user facility reported that smoke was coming from a fresenius dry acid mixer.The machine was removed from service.The customer requested an onsite evaluation by a fresenius field service technician (fst) however this request was cancelled the following day.There was no reported harm to any patients or individuals because of this malfunction.Additional information was requested, however a response has not been received.No parts were returned to the manufacturer for physical evaluation.
 
Additional Manufacturer Narrative
Additional information: d9, h3 plant investigation: the fill valve and display board were returned to the manufacturer for physical evaluation.An external inspection of the fill valve revealed a connector with signs of thermal decomposition.An internal inspection revealed a corroded housing that caused the valve to be stuck.An external inspection of the display board revealed no discrepancies.The display board was connected to a test dry acid dissolution unit, and the display worked as intended.The manufacturer investigation confirmed the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRY ACID 99 GALLON UNIT, ASSY TO SHIP
Common Device Name
DIALYSATE CONCENTRATE FOR HEMODIALYSIS (LIQUID OR POWDER)
Manufacturer (Section D)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key15200827
Report Number0002937457-2022-01332
Device Sequence Number6242200
Product Code KPO
UDI-Device Identifier00840861101009
UDI-Public00840861101009
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
K131611
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 02/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number160148
Device Catalogue NumberRTL160148
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/28/2022
Supplement Date Received by Manufacturer08/11/2022
02/10/2023
02/10/2023
Initial Report FDA Received Date08/10/2022
Supplement Report FDA Received Date08/23/2022
02/10/2023
02/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/24/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-