| Model Number |
160148 |
| Medical Device Problem Code |
Smoking (1585)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A biomedical technician (biomed) at a user facility reported that smoke was coming from a fresenius dry acid mixer.The machine was removed from service.The customer requested an onsite evaluation by a fresenius field service technician (fst) however this request was cancelled the following day.There was no reported harm to any patients or individuals because of this malfunction.Additional information was requested, however a response has not been received.No parts were returned to the manufacturer for physical evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
|
| |
|
Additional Manufacturer Narrative
|
|
Plant investigation: no parts were returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation was performed by a fresenius field service technician (fst).A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The investigation y the manufacturer did not determines that there was a causal relationship between the objective evidence and the alleged event; the alleged event is unconfirmed.
|
| |
|
Event or Problem Description
|
|
A biomedical technician (biomed) at a user facility reported that smoke was coming from a fresenius dry acid mixer.The machine was removed from service.The customer requested an onsite evaluation by a fresenius field service technician (fst) however this request was cancelled the following day.There was no reported harm to any patients or individuals because of this malfunction.Additional information was requested, however a response has not been received.No parts were returned to the manufacturer for physical evaluation.
|
| |
|
Event or Problem Description
|
|
A biomedical technician (biomed) at a user facility reported that smoke was coming from a fresenius dry acid mixer.The machine was removed from service.The customer requested an onsite evaluation by a fresenius field service technician (fst) however this request was cancelled the following day.There was no reported harm to any patients or individuals because of this malfunction.Additional information was requested, however a response has not been received.No parts were returned to the manufacturer for physical evaluation.
|
| |
|
Event or Problem Description
|
|
A biomedical technician (biomed) at a user facility reported that smoke was coming from a fresenius dry acid mixer.The machine was removed from service.The customer requested an onsite evaluation by a fresenius field service technician (fst) however this request was cancelled the following day.There was no reported harm to any patients or individuals because of this malfunction.Additional information was requested, however a response has not been received.No parts were returned to the manufacturer for physical evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, h3 plant investigation: the fill valve and display board were returned to the manufacturer for physical evaluation.An external inspection of the fill valve revealed a connector with signs of thermal decomposition.An internal inspection revealed a corroded housing that caused the valve to be stuck.An external inspection of the display board revealed no discrepancies.The display board was connected to a test dry acid dissolution unit, and the display worked as intended.The manufacturer investigation confirmed the reported issue.
|
| |
|
Search Alerts/Recalls
|