Catalog Number 8065751795
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Conjunctivitis (1784); Endophthalmitis (1835); Uveitis (2122); Blurred Vision (2137); Toxic Anterior Segment Syndrome (TASS) (4469); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 07/27/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that after cataract surgery by using an irrigation/aspiration (i/a) handpiece and phacoemulsification handpiece, a patient experienced severe endophthalmitis or toxic anterior segment syndrome (tass).Additional related information was requested but has not been provided to date.
 
Additional Manufacturer Narrative
The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
New information received indicating that a patient underwent cataract surgery with iol implantation in left eye with medical history of glaucoma.The patient experienced with endophthalmitis after one day of severe cataract surgery and also presented with conjunctival inflammation: 2+, cells in the anterior chamber: 3+.Patient was admitted to the emergency room with symptoms of blurred vision, and it was treated with medications antibiotic+corticosteroid 6 times a day + fluoroquinolones + antibiotics 6 times a day + antibiotics 2 times a day and also treated with intravenous injection such as intravenous injection of antibiotics + glucocorticoid injection + intravitreous injection of glucocorticoid + steroid in subconjunctival form.Event leads to change in visual acuity that reduced to light perception.Current health status of the patient disease was in process of resolution.
 
Additional Manufacturer Narrative
Additional information has been provided in section b.5, b.6 and d.10.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
New information received indicating patient symptoms were resolved and viscoelastic and procedure pak also contributed to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAFLOW II I/A HANDPIECE
Common Device Name
CATHETER, IRRIGATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer (Section G)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15201021
Report Number2523835-2022-00294
Device Sequence Number1856325
Product Code GBX
UDI-Device Identifier00380657517954
UDI-Public00380657517954
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K910245
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 01/11/2023
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number8065751795
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/29/2022
Supplement Date Received by Manufacturer08/03/2022
09/02/2022
12/13/2022
Initial Report FDA Received Date08/10/2022
Supplement Report FDA Received Date08/30/2022
09/29/2022
01/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ADRENALINE; BETADINE 10%; BSS + ADRENALINE; CELESTENE IN SUBCONJUNCTIVAL; CENTURION VISION SYSTEM; CENTURION VISION SYSTEM; CENTURION VISION SYSTEM, ACCESSORY, HANDPIECE; DUOVISC; DUOVISC VISCOELASTIC SYSTEM; INJECTOR MEDJET MB1; INSTRUMENTATION BOX #1; INTRAVITREOUS INJ. OF DEXAMETHASONE; IV INJ. OF ANTIBIOTICS; MEDICONTOUR 640 ADY; MYDRIASERT; OCULAR 5%; PROCEDURE PAK; QUINOFREE; SOLUMEDROL BOLUS INJECTION; STERDEX; STERITRIP; TAVANIC 500MG; TOBRADEX; XYLOCAINE GEL
Outcome Attributed to Adverse Event Other; Hospitalization; Required Intervention;
Patient Age74 YR
Patient SexMale
-
-