MICRO THERAPEUTICS, INC. DBA EV3 AXIUM PRIME BRPL 3D; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
|
Back to Search Results |
|
| Model Number |
APB-1-3-3D-ES |
| Medical Device Problem Codes |
Product Quality Problem (1506); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Event or Problem Description
|
|
Medtronic received a report that the axium prime coil prematurely detached in the sheath.The patient was undergoing treatment for an unruptured, saccular aneurysm located in the right posterior communicating artery.The max diameter was 3mm, and the neck diameter was 1mm.The patient's blood flow was normal, and their vessel tortuosity was moderate.It was reported that the coil was removed, and no additional medical/surgical intervention was required.The pushwire was not bent or broken.There was no friction/difficulty during delivery, the physician did not reposition the coil, no detachment attempts had been made, and the physician did not rotate the delivery pusher.A continuous flush had been administered.It was observed that the use before date of the coil was on (b)(6) 2022, which was prior to the procedure. the coil was replaced, and the patient did not experience any injury.The devices were prepared according to the instructions for use (ifu).Ancillary devices include an envoy 6f guide catheter and a headway 17 microcatheter.
|
| |
|
Event or Problem Description
|
|
Additional information received reported the actual date of the event was 14-may-2022.The device was not used after the use-by date.The cause of the event was not determined.Reporting of the event to medtronic was delayed due to change of the staff in charge.Based on the new information received, the product was not expired at time of use.The device came from consignment stock.
|
| |
|
Additional Manufacturer Narrative
|
|
B3.Event date corrected (14-may-2022).B5.Updated with additional information received.H3.Product analysis: ¿ equipment used: vis (m-81805), 203cm ruler (m-83360) ¿ as found condition: the axium prime coil was returned for analysis within a shipping box and within a plastic bio-pouch.¿ visual inspection/damage location details: the axium prime coil was returned within its introducer sheath with the wavelock at the proximal end.The axium prime pusher actuator interface (ai) and hypotube break indicator (hbi) were found intact.The axium prime pusher was found bent at ~19.5cm and ~150.0cm from the pusher proximal end.The implant coil was found still attached to the pusher within the introducer sheath.The implant coil was found stretched within the polypropylene filament broken.The implant coil distal tip was extending out from the introducer sheath.The implant coil was found damaged and knotted.The axium prime coil could not be pushed or pulled out from within the introducer sheath.¿ testing/analysis: none ¿ conclusion: based on the device analysis and reported information, the customer¿s ¿premature detachment¿ report could not be confirmed as the implant coil was found still attached to the pusher.Regarding the damage (stretching) found with the returned axium prime implant coil it is likely the implant coil became knotted subsequently stretching during resheathing.The implant coil can become knotted during handling or preparation.H6.Coding updated based on analysis results.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|