STRYKER ENDOSCOPY-SAN JOSE OMEGA 4.75MM PEEK KNOTLESS ANCHOR SYSTEM, SINGLE; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number 3910-500-471
Health Effect - Clinical Codes Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/15/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that the screw broke during the procedure.The broken screw remains in the patient's bone.
 
Event or Problem Description
It was reported that the screw broke during the procedure.The broken screw remains in the patient's bone.
 
Additional Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: screw broken.Probable root cause: design: inadequate raw material selection for screw.Design geometry too sensitive to withstand clinical impaction loads.Screw/driver interface not designed to maintain proper assembly.Process: screw or driver not manufactured to specification.Incorrect material used during manufacturing.Screw and driver not assembled to specification.Application: excessive force or leveraging of the screw against the bone.Inadequate assessment of bone quality, pilot hole not prepared.The reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.H3 other text : 81.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMEGA 4.75MM PEEK KNOTLESS ANCHOR SYSTEM, SINGLE
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
andrea zenere
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key15201169
Report Number0002936485-2022-00452
Device Sequence Number4800305
Product Code MBI
UDI-Device Identifier37613327464819
UDI-Public37613327464819
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K181083
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number3910-500-471
Device Lot Number21306AG2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/16/2022
Supplement Date Received by Manufacturer07/16/2022
Initial Report FDA Received Date08/10/2022
Supplement Report FDA Received Date09/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-