| Model Number |
1181200777T |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/04/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Event or Problem Description
|
|
The customer reported that the syringe tip breaks off in the intravenous (iv) connecting tubing.There was no patient harm reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Since a lot number was not provided, the device history record review could not be performed.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.A sample was not received from the customer for the investigation.Because a sample was not returned, we were unable to perform a follow up investigation to include functional and visual evaluations to confirm the issue and root cause analysis.The retained samples by the manufacturer were visually checked.There were no deviations or issues in any components.At this time, a corrective and preventive action is not deemed necessary.We will keep monitoring the process for any adverse trends that require immediate attention.This complaint will be used for qa tracking and trending purposes.
|
| |
|