• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 AXIUM PGLA 3D; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 AXIUM PGLA 3D; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number PC-2-6-3D
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/14/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The event is reported at this time based on analysis results which indicate a reportable event.Product analysis:¿ equipment used: video inspection system (m-78210) with microscope, ruler (m-83361), camera (panasonic lumix dmc-zs5) ¿ as found condition: the axium coil was returned for evaluation within the introducer sheath and inner pouch; inside of a biohazard bag and a shipping box.¿ visual inspection/damage location details: the implant coil appeared to be detached from the pushwire.The implant coil was found to be damaged and stretched with the polypropylene filament intact.The shield coil was intact with no damage.The coin was not present against the lumen stop as it was pulling back.The pusher was found broken at the distal break indicator (manual detachment location); it appeared that the manual detachment was attempted at this location.The ai and coupler tubing were present and intact; it appeared that the mechanical detachment was not attempted at this location.Bends were found along the length of the pushwire.All other subassemblies appeared to be normal, and no other anomalies were observed.¿ testing/analysis: the axium coil was removed from the sheath with difficulty.The ¿wave lock¿ portion of the introducer sheath was present near the proximal end.The implant coil tip od (outer diameter) was measured to be 0.0114¿ which is within specifications.The introducer sheath id (inner diameter) was measured to be 0.0190" (0.44mm) found to be within specifications.The detach element was in good shape and the detachment ball was measured to be 0.0038" and found to be within specifications.Under the microscope, the outer jacket was then removed to gain access the coin.The coin was measured in 3 locations (0.072mm @ 0.063mm; measured 0.083mm @ 0.127mm; measured 0.099mm @ 0.275mm) and found to be within specifications.The inner diameter of the lumen stop, and the inner diameter of the retainer ring were found to be visually acceptable.The lumen stop inner diameter (id) was measured to be 0.00265¿ and found to be within specification.The retainer ring inner diameter (id) was measured to be 0.00459¿ and found to be within specifi cation.¿ conclusion: based on the analysis performed, the customer¿s report of ¿coil resistance/stuck in sheath¿ was confirmed as the implant coil was found stuck inside the introducer sheath.The returned axium coil was found to be damaged and stretched.The pushwire was damaged at several locations.However, the cause root cause and the cause for damages could not be determined.The customer did not report any issues pushing the implant coil out from the introducer sheath during the hydration process.Therefore, it is possible that damages occurred during preparation or due to an improper hubbing technique (over tightening of the rhv).In addition, the implant coil also found detached from the pushwire.Additionally, the pushwire was found broken at the manual detachment location with the coin pulled back from the lumen stop.Pulling back of the coin from the lumen stop may have contributed to the detachment of the implant coil from the pushwire.There was no non-conformance to specification identified that led to the reported issue.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
Medtronic received information regarding an axium coil that had resistance in the sheath.The patient was undergoing a coil embolization procedure to treat a dural arteriovenous fistula.Blood flow and vessel tortuosity were normal.It was reported that the axium coil and all accessory devices were prepared as indicated in the instructions for use (ifu).The coil could not be advanced out of the sheath into the rhv/microcatheter and was stuck.The coil was removed and tried in vitro on the table but still could not be pushed out of the introducer sheath.There was no harm or injury to the patient.Ancillary devices: 6f sheath, 6f neuron max 088 sheath, 5f sophia guide catheter, headway duo microcatheter additional information received reported that the coil was prepared like described above (as per ifu).When the coil was inserted into the microcatheter¿s rhv it would not advance in the introducer sheath and was getting stuck to where only the coil could be pushed out but it wouldn¿t advance any further.Coil was removed and they tried advancing it on the table to see what the issue was and same thing happened.Looked as if it was getting stuck at the proximal end of the introducer sheath and pusher wire.Coil didn¿t enter patient and was removed from table.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIUM PGLA 3D
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key15201271
Report Number2029214-2022-01331
Device Sequence Number15056128
Product Code KRD
UDI-Device Identifier00847536033184
UDI-Public00847536033184
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K081465
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/16/2024
Device Model NumberPC-2-6-3D
Device Catalogue NumberPC-2-6-3D
Device Lot NumberB222529
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/23/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/22/2022
Initial Report FDA Received Date08/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/17/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age38 YR
Patient SexFemale
-
-