• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 AXIUM 3D; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 AXIUM 3D; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number QC-14-40-3D
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product analysis: ¿ as found condition: the pushwire was returned for evaluation within the introducer sheath; inside of a biohazard bag and a shipping box.There was no implant coil returned with the pushwire.¿ visual inspection/damage location details: the implant coil appeared to be detached from the pushwire.The shield coil was intact with no damage.The coin was not present against the lumen stop as it was pulling back.The pusher was found broken at the distal break indicator (manual detachment location); it appeared that the manual detachment was attempted at this location.The ai and coupler tubing were present and intact; it appeared that the mechanical detachment was not attempted at this location.Bends were found at 24.0cm to 37.0cm from the proximal end.All other subassemblies appeared to be normal, and no other anomalies were observed.¿ testing/analysis: the pushwire was removed from the sheath with difficulty.The ¿wave lock¿ portion of the introducer sheath was present near the proximal end.The introducer sheath id (inner diameter) was measured to be 0.0190" (0.44mm) found to be within specifications.Under the microscope, the outer jacket was then removed to gain access the coin.The coin was measured in 3 locations (0.075mm @ 0.063mm; measured 0.086mm @ 0.127mm; measured 0.094mm @ 0.275mm) and found to be within specifications.The inner diameter of the lumen stop, and the inner diameter of the retainer ring were found to be visually acceptable.The lumen stop inner diameter (id) was measured to be 0.00270¿ and found to be within specification.The retainer ring inner diameter (id) was measured to be 0.00460¿ and found to be within specification.¿ conclusion: based on the analysis performed, the customer complaint was confirmed as the pushwire was returned stuck inside the introducer sheath.In addition, the pushwire was found damaged.However, the root cause and the cause for damages could not be determined.The customer did not report any issues pushing the implant coil out from the introducer sheath during the hydration process.Therefore, it is possible that damages occurred during preparation or due to an improper hubbing technique (over tightening of the rhv).In additionally, the pushwire was found broken at the manual detachment location with the coin pulled back from the lumen stop.Pulling back of the coin from the lumen stop may have contributed to the detachment of the implant coil from the pushwire.There was no non-conformance to specification identified that that led to the reported issue.Since the implant coil was not returned; any contribution of the implant coil to the issues could not be determined.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received a report that the axium coil was not able to enter the catheter due to severe resistance in the sheath. the coil was replaced, and the patient did not experience any injury.The devices were prepared according to the instructions for use (ifu).The patient was undergoing treatment for an unruptured, saccular aneurysm located in the right posterior communicating artery.The max diameter was 15mm, and the neck diameter was 8mm.The patient's vessel tortuosity was normal, and their vessel tortuosity was moderate.Ancillary devices include an envoy 6f guide catheter and headway 17 microcatheter.Additional information received reported that during preparation, the introducer sheath was held vertically when the implant coil was pulled back inside during hydration.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIUM 3D
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key15202108
Report Number2029214-2022-01327
Device Sequence Number6413210
Product Code KRD
UDI-Device Identifier00847536030176
UDI-Public00847536030176
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K081465
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/02/2023
Device Model NumberQC-14-40-3D
Device Catalogue NumberQC-14-40-3D
Device Lot NumberB125791
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/23/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/22/2022
Initial Report FDA Received Date08/10/2022
Date Device Manufactured12/02/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-