MEDTRONIC SOFAMOR DANEK USA, INC SOLERA VOYAGER 4.75; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
|
Back to Search Results |
|
Model Number 6640008 |
Device Problem
Break (1069)
|
Patient Problem
Device Embedded In Tissue or Plaque (3165)
|
Event Date 07/13/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a healthcare provider (hcp) via manufacturer representative regarding a spinal product used in unknown spinal therapy.It was reported that during the first screw implantation procedure (l4), the screwdriver was broken and stuck into screw head and cannot be taken out.When another screw was implanted, the other driver was broken.No further complications or symptoms were reported.Additional information was received via manufacturer representative that there was no malfunction associated with the screws.Broken fragments of drivers was left inside the patient as surgeon failed to pull out the screws due to patient's high bone density.There is no revision surgery was scheduled or planned as the screw was too hard to unscrew during the surgery, and the mri/ct image showed the screw and rod were stable (2 day after surgery).
|
|
Manufacturer Narrative
|
Product analysis of part # 6640008, lot # nm19a029 visual inspection confirmed the entire torx tip of instrument has been sheared off and the attachment pin to the internal bushing has been broken off.Optical examination of the fracture surface revealed a fairly flat fracture surface and circular material flow.This type of damage is consistent with torsional overload.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|