|
Device evaluation summary: medtronic conducted an investigation based upon all information received.It was reported that this ht70 ventilator would not power on.The service personnel (sp) inspected the ventilator and confirmed the reported issue.The sp replaced the control board and power switch.The ventilator passed all the tests and calibrations as per manufacturer's specifications at the time of service.One power switch was returned to medtronic for failure investigation.The returned power switch was visually inspected and functionally tested with no malfunction or product deficiency observed.One control board was returned to medtronic for failure investigation.Visual inspection showed no anomalies.The returned control board was installed in test ventilator and found the unit was unable to power up.Further inspection of the control board found capacitor c92 faulty.The analysis determined that the failure of capacitor c92 while the unit is on would lead to a loss of power/ventilation.There is an existing internal investigation to address this issue.The cause of the event was isolated to faulty capacitor c92 of the control board.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it has met all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|