| Model Number |
M00513750 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The complainant indicated that the ultraflex esophageal stent will not be returned for evaluation; therefore a device analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation on (b)(6) 2022 that an ultraflex esophageal ng distal release covered stent was implanted during a procedure performed on an unknown date.On (b)(6) 2022, post stent placement, it was noted that the ultraflex esophageal stent had a partial, distal migration.There were no patient complications reported due to this event.Note: no further information has been obtained despite good faith efforts.
|
| |
|
Additional Manufacturer Narrative
|
|
Blocks a2 (age at time of event), a3, a4, b1, b2 (outcomes attrib to adv event), b5, d6a, d7b, e1 (initial reporter facility name, address 1, city, country, zip/post code, phone number), e3 (occupation), g2 (report source, report source other), h1, h6 (impact codes), h8 and h10 have been updated with additional information received august 23, 2022 and august 26, 2022.Block h6: medical device problem code a010402 captures the reportable event of ultraflex esophageal stent migration.Impact code f23 is being used to capture the additional intervention of stent repositioning and removal of the migrated ultraflex esophageal stent.Block h10: the complainant indicated that the ultraflex esophageal stent was disposed and will not be returned for evaluation; therefore a device analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation on july 25, 2022 that an ultraflex esophageal ng distal release covered stent was implanted during a procedure performed on an unknown date.On july 01, 2022, post stent placement, it was noted that the ultraflex esophageal stent had a partial, distal migration.There are no known patient complications reported due to this event.Additional information received on august 23, 2022 and august 26, 2022.Note: this report pertains to one of two devices used on the same patient.Refer to manufacturer report and 3005099803-2022-04526 for the associated device information.The ultraflex esophageal stent was implanted in the esophagus to treat a 10cm malignant stricture during an esophagogastroduodenoscopy (egd) with stent placement procedure performed on june 30, 2022.The patient's anatomy was tortuous and was not dilated prior to stent placement.On july 01, 2022, post stent placement, the ultraflex esophageal stent (the subject of this report) had a partial, distal migration and was repositioned with forceps.The ultraflex esophageal stent remained implanted and the procedure was completed.On july 11, 2022, egd procedure was performed and the ultraflex esophageal stent was removed from the patient with forceps.A second ultraflex esophageal stent (the subject of mfr.Report # 3005099803-2022-04526) was to be implanted; however, "the stent had only opened partially".The second ultraflex esophageal stent was removed from the patient and the procedure was completed with another ultraflex esophageal stent.
|
| |
|
Additional Manufacturer Narrative
|
|
Block b5 has been updated with additional information received on august 30, 2022.Block h6: medical device problem code a010402 captures the reportable event of ultraflex esophageal stent migration.Impact code f23 is being used to capture the additional intervention of stent repositioning and removal of the migrated ultraflex esophageal stent.Block h10: the complainant indicated that the ultraflex esophageal stent was disposed and will not be returned for evaluation; therefore a device analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation on july 25, 2022 that an ultraflex esophageal ng distal release covered stent was implanted during a procedure performed on an unknown date.On (b)(6) 2022, post stent placement, it was noted that the ultraflex esophageal stent had a partial, distal migration.There are no known patient complications reported due to this event.**additional information received on august 23, 2022 and august 26, 2022** note: this report pertains to one of two devices used on the same patient.Refer to manufacturer report # (b)(4) and (b)(4) for the associated device information.The ultraflex esophageal stent was implanted in the esophagus to treat a 10cm malignant stricture during an esophagogastroduodenoscopy (egd) with stent placement procedure performed on (b)(6).The patient's anatomy was tortuous and was not dilated prior to stent placement.On (b)(6) 2022, post stent placement, the ultraflex esophageal stent (the subject of this report) had a partial, distal migration and was repositioned with forceps.The ultraflex esophageal stent remained implanted and the procedure was completed.On (b)(6) 2022, egd procedure was performed and the ultraflex esophageal stent was removed from the patient with forceps.A second ultraflex esophageal stent (the subject of mfr.Report # (b)(4)) was to be implanted; however, "the stent had only opened partially".The second ultraflex esophageal stent was removed from the patient and the procedure was completed with another ultraflex esophageal stent.**additional information received on august 30, 2022** the previously implanted ultraflex esophageal stent (the subject of this report) had slipped up from the target anatomy.A second ultraflex esophageal stent (the subject of mfr.Report # (b)(4)) was deployed and positioned below the previously implanted ultraflex esophageal stent; however, the second ultraflex esophageal stent (the subject of mfr.Report # (b)(4)) did not fully expand (partially opened).Both ultraflex esophgaeal stents were removed from the patient using grasping forceps.
|
| |
|