| Model Number |
393012 |
| Medical Device Problem Code |
Premature End-of-Life Indicator (1480)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This device was explanted due to eri with unexpected battery behavior.No adverse patient events were reported.Should additional information become available, this file will be updated.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is currently not available for analysis.No conclusion can be drawn at this time.No additional information is available at the moment.The file is closed.The investigation will be re-opened should additional data become available.
|
| |
|
Additional Manufacturer Narrative
|
|
Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this icd were re-investigated.All production steps were performed accordingly, and the final acceptance test proved the device functions to be as specified.Next, the device was interrogated.The interrogation could be properly performed and revealed the mol1 battery status.Battery trend data showed a temporary voltage drop followed by recovery of the battery voltage.An unexpected battery voltage trend could be confirmed during analysis.Please note, that this icd is subjected to the field safety corrective action, bio-lqc, initiated in march 2021.
|
| |
|
Search Alerts/Recalls
|