| Model Number |
801764 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The reported complaint was confirmed.The srt replaced the solid state relay.Safety/verification testing was completed successfully.The unit operated to the manufacturer's specifications.This complaint is related to (b)(4)/mfr #1828100-2022-00283.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
The service repair technician (srt) reported that during testing of the device at the service center, the heart lung machine (hlm) solid state relay failed.There was no patient involvement.
|
| |
|