• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THORATEC CORPORATION HEARTMATE SYSTEM MONITOR TO POWER MODULE CABLE; VENTRICULAR (ASSIST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THORATEC CORPORATION HEARTMATE SYSTEM MONITOR TO POWER MODULE CABLE; VENTRICULAR (ASSIST) BYPASS Back to Search Results
Model Number 103859
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/18/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
There was no patient involved in this event.The pma# provided is associated with most recent approval.A supplemental report will be submitted once the manufacturer¿s investigation is completed.Initial reporter phone number: (b)(6).
 
Event or Problem Description
It was reported that the system monitor to power module cable was damaged and the wires were exposed next to the connector.The system monitor was still fully functional, but the hospital wished to exchange the system monitor to prevent further technical issues.
 
Additional Manufacturer Narrative
Manufacturer's investigation conclusion: the reported event of damage to heartmate system monitor to power module cable was confirmed via review of the submitted photo.This photo revealed that the cable had slightly pulled out of its strain relief at one end, exposing its inner wires and shielding.No damages to the inner wires were conclusively observed, and no functional issues regarding the cable were reported.The cable was not returned for evaluation, and the root cause of the reported event was unable to be conclusively determined via this analysis.The heartmate 3 instructions for use section 4 entitled ¿system monitor¿ provides instruction on how to properly install and uninstall the system monitor data cable.The heartmate system monitor instructions for use inform the user to refer any servicing of heartmate lvas equipment to trained service personnel only.The heartmate 3 instructions for use section 8 entitled ¿equipment storage and care¿ describes how to properly store, transport, and clean all heartmate equipment.The heartmate 3 patient handbook cautions users to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.The device history records for the system monitor data cable were unable to be reviewed, as they reside with the original equipment manufacturer (oem).The lot number of the cable was also not provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTMATE SYSTEM MONITOR TO POWER MODULE CABLE
Common Device Name
VENTRICULAR (ASSIST) BYPASS
Manufacturer (Section D)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer (Section G)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer Contact
bob fryc
6035 stoneridge drive
pleasanton, CA 94588
7818528204
MDR Report Key15203655
Report Number2916596-2022-12863
Device Sequence Number14632028
Product Code DSQ
UDI-Device Identifier00813024010753
UDI-Public813024010753
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P160054
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number103859
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/18/2022
Supplement Date Received by Manufacturer10/04/2022
Initial Report FDA Received Date08/10/2022
Supplement Report FDA Received Date10/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-