• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH MAGELLAN INSULIN 1ML 29GX1/2; NEEDLE, HYPODERMIC, SINGLE LUMEN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH MAGELLAN INSULIN 1ML 29GX1/2; NEEDLE, HYPODERMIC, SINGLE LUMEN Back to Search Results
Model Number 8881892910
Medical Device Problem Code Fail-Safe Problem (2936)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event or Problem Description
The customer reported that when the nurse pulling out the orange cap and the safety guide was also pulling out making the syringe not usable.Additional information was received and stated that the orange cap was not hard to pull.Per customer, the entire safety mechanism was sticking out above white safety guard when pulling out the cap.There was no patient involved and no injuries reported.The issue was discovered before use.
 
Additional Manufacturer Narrative
See section d4 for the expiration date provided, as the sample returned by the customer has expiration date on it.The lot number was provided, and the device history record (dhr) was reviewed indicating that the product was released accomplishing all quality standards.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.The product was manufactured and packaged in february 2022.The device was received for evaluation and the reported condition has been confirmed.A corrective and preventive action has been initiated to address the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGELLAN INSULIN 1ML 29GX1/2
Common Device Name
NEEDLE, HYPODERMIC, SINGLE LUMEN
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
1222 sherwood rd
norfolk NE 68701
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key15203980
Report Number1915484-2022-01287
Device Sequence Number16661134
Product Code FMI
UDI-Device Identifier10884521005938
UDI-Public10884521005938
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8881892910
Device Catalogue Number8881892910
Device Lot Number203816X
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/03/2022
Supplement Date Received by Manufacturer08/03/2022
Initial Report FDA Received Date08/10/2022
Supplement Report FDA Received Date09/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-