| Model Number |
ZZ19KN70ZZZ006N |
| Medical Device Problem Code |
Loss of or Failure to Bond (1068)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/03/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation: two tracheostomy tubes were returned for evaluation.The first sample was visually inspected and manipulated to search for any gaps or holes, and none were found.The second sample was also visually inspected and manipulated.The neck flange was adhered to the device, although a portion was found to not be fully flush with the neck flange section when manipulated.This confirms the reported event.A review of manufacturing records found no discrepancies or anomalies, and no trends relevant to this product line and complaint category were found during review of complaint data.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to use, it was noticed that the neck flange is not properly glued and gaps are able to be visualized.Customer reported that the gaps seem to be filled with "superslick".No patient involvement.
|
| |
|