Event or Problem Description
|
|
It was reported when using the bd vacutainer® lithium heparinn (lh) blood collection tubes, the device experienced underfill or low draw of a tube with blood.The following information was provided by the initial reporter.The customer stated: it was reported by the customer that tubes are experiencing vacuum issues.Barbara carson : (b)(6) 2022 15:34:04 (gmt) sm 367878, lot #1137406 vacuum issues.
|
| |
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
H.6.Investigation summary: "material #: 367878.Lot/batch #: 1137406.Bd had not received samples or photos for investigation.Therefore, ten (10) retention samples from bd inventory were evaluated by functional testing and no issues were observed relating to underfill as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode of underfill.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.".
|
| |
|
Event or Problem Description
|
|
It was reported when using the bd vacutainer® lithium heparinn (lh) blood collection tubes, the device experienced underfill or low draw of a tube with blood.The following information was provided by the initial reporter.The customer stated: it was reported by the customer that tubes are experiencing vacuum issues.(b)(6) 2022-07-27 15:34:04 (gmt).Sm 367878, lot #1137406 vacuum issues.
|
| |
|