Model Number MMT-1715KM
Medical Device Problem Code Battery Problem (2885)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/08/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Insulin pump trace download analysis confirmed the insulin pump alarmed multiple low battery alerts.The power management tool confirmed low battery alarms were triggered when backup battery loaded voltage was less that 3.5 volt for four consecutive hours due to resistance issue at electrical board.After disconnecting and reconnecting the internal battery connector on electrical board, insulin pump was monitored and functioned properly.Device received with minor scratched lcd window, cracked retainer and scratched case.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
Information received by medtronic indicated that the insulin pump had a low battery alert.Customer stated that battery did not last more than 1 week.No harm requiring medical intervention was reported.The device was returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUMP MMT-1715KM 630G 3ML BLACK MEDI  
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*  00777-3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*   00777-3869
Manufacturer Contact
tricha miles
ceiba norte ind. park #50 road
juncos 00777--386
*   00777-3869
7635140379
MDR Report Key15207859
Report Number2032227-2022-309136
Device Sequence Number4800052
Product Code OZO
UDI-Device Identifier00643169873834
UDI-Public(01)00643169873834
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-1715KM
Device Catalogue NumberMMT-1715KM
Device Lot NumberHG1L1QB
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/19/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/06/2022
Initial Report FDA Received Date08/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/13/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age51 YR
Patient SexMale
Patient Weight100 KG
-
-