| Model Number |
SRL-380 |
| Medical Device Problem Codes |
Device Appears to Trigger Rejection (1524); Material Rupture (1546)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
07/13/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: analysis of the returned device identified: creases fold, wear abrasion and weight to the spec.Bubbles and voids in the gel observed after autoclave cycle.The unit is not ruptured.Based on the device analysis the final assessment is no issues found related with the manufacturing process.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iii and rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side deflation and capsular contracture, baker grade iii.Device has been explanted.
|
| |
|