|
Catalog Number UNKNOWN |
Device Problems
Structural Problem (2506); Device Tipped Over (2589); Detachment of Device or Device Component (2907); Unintended Movement (3026)
|
Patient Problems
Pain (1994); Perforation of Vessels (2135); Depression (2361); Peripheral Edema (4578)
|
Event Type
Injury
|
Manufacturer Narrative
|
Occupation: non-healthcare professional.Investigation: the reported allegations have been investigated based on the information provided to date.The following allegations have been investigated: fracture, vena cava perforation, and migration.Filter fracture has been reported and may be either symptomatic or asymptomatic.Fracture of a filter leg may be due to repetitive motion on a filter leg in an unusual, stressed position, such as a filter leg penetrating/perforating the ivc; or a filter leg being caught in a side branch (e.G., a renal vein).Other potential causes of filter fracture may include excessive force or manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Retrieval of a fractured filter or filter fragments (including embolized fragments) using endovascular techniques has been reported.Potential adverse events that may occur include, but are not limited to, the following: filter fracture, filter or filter fragment embolization, trauma to adjacent structures.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: vascular trauma, vena cava perforation, vena cava penetration.Filter or filter fragment migration and (or) embolization (e.G., movement to the heart or lungs) has been reported.Filter or filter fragment movement has occurred in both the cranial and caudal direction and may be either symptomatic or asymptomatic.Potential causes may include filter placement in ivcs with diameters smaller or larger than those specified in these instructions for use; improper deployment; deployment into thrombus; dislodgement due to large thrombus burdens; and (or) excessive force or manipulations near an in situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: filter migration, trauma to adjacent structures.Catalog number and lot number are unknown, however, the alleged tulip is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications, and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
The following information is alleged: the patient received a gunther tulip inferior vena cava (ivc) filter on an unknown date approximately in 2007.Approximately 14 years later, the patient underwent a computed tomography (ct) scan which revealed that the filter had perforated the ivc wall.The scan further revealed that the filter was fractured as at least one strut had migrated to the right ventricle.Hospital and medical records have been requested, but not yet provided.
|
|
Manufacturer Narrative
|
Investigation: investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.The following allegations have been investigated: tilt, pain, lower extremity swelling, limited mobility.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/perforation; and (or) result in loss of filter efficiency.Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt.Unknown if the reported pain, lower extremity swelling, limited mobility is directly related to the filter and unable to identify a corresponding failure mode at this point in time.Catalog and lot numbers are unknown, however, the alleged tulip is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications, and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
Additional information provided patient is alleging device migration, tilt, and fracture, vena cava perforation.Patient notes and further alleges experiencing swelling and pain in the left leg resulting in difficulty walking and requiring pain medication ; soreness and aching in feet inhibiting "any kind of activity"; depression.Per a ct abdomen: "vascular: the proximal tip of the ivc filter is approximately 1.5 cm distal to the origin of the left renal vein.The apex of the filter abuts the posterior wall of the ivc.There are several struts that extend beyond the confines of the ivc up to 6 mm (series 2, image 77).A medial strut abuts the aorta without definitive perforation or penetration.There is an ill-defined linear high density focus within the region of the right ventricle of uncertain etiology; however, migrated strut cannot be excluded".Successful ivc filter retrieval: "findings: 1.No ivc thrombus/anomaly 2.Successful retrieval of a gunther-tulip ivc filter.The filter was completely intact and without defect".
|
|
Search Alerts/Recalls
|
|
|