| Model Number |
N/A |
| Medical Device Problem Codes |
Positioning Failure (1158); Peeled/Delaminated (1454)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported surgeon went to impact liner and liner would not seat the shell.Liner was removed and plastic from locking mechanism was peeling off.Attempts have been made and additional information on the reported event is unavailable at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Concomitant medical products: 010000666- g7 pps ltd acet shell 58g- 7167172.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection shows the lot number match complaint and item number is not etched on part.There is damage to face of the part.There is raised material on a couple areas between the scallops.There are 2 scallops that are raised above the flat face of the part, which should be flush.The rib shows damage under magnification but no photos were taken.No peeling was observed.A review of the device history records identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No additional information on the reported event.
|
| |
|