• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: O&M HALYARD, INC. FLUIDSHIELD* 1 PROCEDURE MASK WITH SO SOFT* LINING AND SO SOFT* EARLOOPS, YELLOW; FLUIDSHIELD* LEVEL 1 PROCEDURE MASK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

O&M HALYARD, INC. FLUIDSHIELD* 1 PROCEDURE MASK WITH SO SOFT* LINING AND SO SOFT* EARLOOPS, YELLOW; FLUIDSHIELD* LEVEL 1 PROCEDURE MASK Back to Search Results
Model Number 25867
Medical Device Problem Code Component Incompatible (1108)
Health Effect - Clinical Code Local Reaction (2035)
Date of Event 03/26/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Patient 1 of 4.Employee experienced reactions on four occasions associated with mask use.The reaction(s) consistent of tongue feeling thick with itchy throat.Initial reaction occurred on (b)(6) 2022 and subsequently on (b)(6) 2022, (b)(6) 2022 to (b)(6) 2022 and (b)(6) 2022 to (b)(6) 2022.Employee health prescribed intramuscular epinephrine, intravenous prednisone, benadryl, pepcid, solumedrol, and hydroxyzine.The employee followed up with an allergist and had two patch tests of steroids.Employee was referred to pulmonologist and rheumatologist.She currently continues benadryl, pepcid, solumedrol, hydroxyzine, 20 mg prednisone, allegra (twice a day), and zyrtec (twice a day) by mouth.
 
Additional Manufacturer Narrative
The product involved in the event reported has not been returned but a representative sample is anticipated.A follow-up report will be provided upon the conclusion of the investigation.All information reasonably known as of august 11, 2022 has been included in this medical device report.The information provided by o&m halyard, inc.Represents all of the known information at this time.O&m halyard, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the o&m halyard, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that a o&m halyard, inc.Product is defective or has caused serious injury.
 
Additional Manufacturer Narrative
Representative product sample was returned to the failure analysis lab for evaluation.The samples were evaluated under ambient light conditions and appeared intact.There are no sharp edges or corners observed on the masks.There are no loose fibers observed on the masks.The samples were inspected and measured by comparing both to specification component.The samples were found to be assembled according to the specification drawing.Device history record (dhr) evaluation was performed for the product code 25867 and lot number am1310061 identified in the complaint.A quality nonconformance was not reported at the time of production.Based on the low to negligible health hazard presented by the components of the level 1 mask in combination with user exposures of less than twenty-four (24) hours and acceptable biocompatibility results, the level 1 masks have a wide margin-of-safety for users when worn as indicated.Manufacturing conducts visual and functional inspection of each device throughout production.According to manufacturing records, the device was manufactured per procedures and specifications, then released by quality assurance.All information reasonably known as of october 13, 2022 has been included in this medical device report.The information provided by o&m halyard, inc.Represents all of the known information at this time.O&m halyard, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the o&m halyard, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that a o&m halyard, inc.Product is defective or has caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
FLUIDSHIELD* LEVEL 1 PROCEDURE MASK
Manufacturer (Section D)
O&M HALYARD, INC.
1 edison drive
alpharetta GA 30005
Manufacturer (Section G)
LA ADA DE ACUÑA
km 4.5 carreterra presa
la amistad
cd. acuña, coahuila mexico 26220
MX   26220
Manufacturer Contact
nichole early
1 edison drive
alpharetta, GA 30005
8287820529
MDR Report Key15211364
Report Number9616096-2022-00005
Device Sequence Number4622672
Product Code FXX
UDI-Device Identifier30680651258671
UDI-Public30680651258671
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K110455
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 09/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number25867
Device Catalogue Number991025867
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/13/2022
Supplement Date Received by Manufacturer09/26/2022
Initial Report FDA Received Date08/11/2022
Supplement Report FDA Received Date10/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-