Model Number SHPX260
Medical Device Problem Codes Material Rupture (1546); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Capsular Contracture (1761); Deformity/ Disfigurement (2360)
Date of Event 02/02/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a african american female who underwent a breast reconstruction revision with a mentor memorygel, 560cc on the left, and 260cc on the right side silicone breast implant experienced bilateral silent rupture postoperative.As a result, patient underwent bilateral replacements as follow: l) cat#:9583050, sn#: (b)(4), r) cat#:7505373; sn: (b)(4) on (b)(6) 2022.This report relates to the right prosthesis.
 
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: silent rupture.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
Based on postmarket study data, the alert/awareness dates for this event have been updated.As a result, the initial report is retroactively considered late.Additional information received on jul 2, 2022 indicated that the patient experienced left-sided anisomastia, and right-sided capsular contracture unknown grade postoperatively.During the explant surgery both devices were found to be intact.Manufacturer¿s reference number: (b)(4) the initial report section b5 mistakenly reported incorrect replacements' cat#s.Correction: replacements as follow: r) cat#: shpx260; sn#: (b)(6), l) cat#: shpx560, sn#: (b)(6).
 
Additional Manufacturer Narrative
On march 22, 2023, per clinician re-review of the event, pec - adverse event was removed, and added pec - rupture symptomatic (post-implant) was added.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
Per additional information received on august 25, 2023, the patient also experienced left side capsular contracture grade iv and right sided grade iii.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
Additional information received on february 08, 2024 indicated that possible bilateral rupture mri (b)(6) 2022, was clarified during surgery doe (b)(6) 2022, both devices were found intact.Manufacturer¿s reference number (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
michelleann garcia
3041 skyway circle north
irving, TX 75038
646591-7981
MDR Report Key15213705
Report Number1645337-2022-09374
Device Sequence Number2199987
Product Code FTR
UDI-Device Identifier00081317025221
UDI-Public00081317025221
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/17/2022
Device Model NumberSHPX260
Device Catalogue NumberSHPX260
Device Lot Number7502195
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/22/2022
Supplement Date Received by Manufacturer07/02/2022
03/22/2023
08/25/2023
02/08/2024
Initial Report FDA Received Date08/11/2022
Supplement Report FDA Received Date10/18/2022
03/24/2023
08/31/2023
02/12/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/18/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient Weight58 KG
Patient EthnicityNon Hispanic
Patient RaceBlack Or African American
-
-