A steris account manager was informed of the event that occurred with the dual articulating headrest assembly on (b)(6) 2022.The account manager went to the user facility to perform the in service for the surgical table and headrest on 7/18/2022.The user facility personnel utilized their 4095 surgical table and dual articulating headrest assembly prior to the scheduled in-service training, which is when the event occurred.The employee subject of the reported event attempted to attach the headrest onto their 4095 surgical table.While attempting to attach the headrest the employee squeezed the left lever of the dual articulating headrest assembly resulting in the employee's fingers being pinched in a pinch point of the device.The steris 4095 general surgical table operator manual safety and precautions section states, "warning - pinching hazard: pinch points are created during tabletop articulation.Become familiar with all pinch points prior to operating the table.To avoid serious injury, keep limbs, fingers, and other body areas clear of all pinch points when positioning the table.When manually positioning table sections, stay clear of potential pinching hazards." "warning - pinching and tipping hazard: patient injury may result if the operator of this table is not completely familiar with the controls for tabletop articulation and tabletop operation." the account manager inspected the headrest and found it to be operating according to specification.The account manager conducted the in-service training on the use and operation of the 4095 surgical table, including use of the dual articulating headrest assembly.No additional issues have been reported.
|